Phased-array
- Customized focusing, electronic scanning, and
sectorial scanning
- Inspection mode: pulse echo with electronic
selection of active channels, DDF
- Fast multiplexing of focal laws during electronic
scanning (stored on 32MB RAM)
Pulsers
- Adjustable voltage: 10 to 80 V with 1V steps
- Negative rectangular pulse, adjustable width: 20 ns
to 1.2 µs, steps of 2.5 ns
- Fall time < 10ns (80V, 50Ω)
- Maximum PRF: 30kHz
Receivers
- Bandwidth: 0.8 to 20 MHz
- Adjustable gain on each channel from 0 to 80 dB
- Adjustable analog DAC 80 dB (max. 20 dB/µs)
synchronized on events
- Crosstalk between two channels: gain > 50 dB
- Maximum input signal amplitude ± 1V
Digitizer
- Maximum sampling frequency: 100 MHz (adjustable from
100 MHz to 6.6 MHz)
- Range: 10 bits
- Input impedance: 50Ω
- Global delay: 0 up to 1.6 ms, steps of 10ns
- Signal max length: 8kB per channel, 50kB on the sum.
- Delay laws at transmission/reception: 0 to 20 µs,
steps of 2.5 ns
Input-Output
- USB2 for data transfer
- 1 FRB (Hypertronix) for phased-array probe
- Encoders input
- External power supply input
|
Dimensions
- Length: 200mm (7.9”)
- Width: 110mm (4.3”)
- Thickness: 60mm (2.4”)
Interface
- Dynamic link libraries (dlls) provided
- Connectors:
- USB2 connector
- Probe connector
- Optional: board delivered in case providing the
listed connections
|